게시글검색

HUMAN MEDITEK CO., LTD.

Innovative Technology for Human Benefits

HMTS-142D

HMTS-142D is a pass-through low temperature hydrogen peroxide sterilizer designed for the Central Sterile Services Departments with a need for an isolated clean room. The automatic sliding door allows for easy loading and increase productivity.


HMTS-142D

Dimensions and Weight

Size
Width 1050 mm
Height* 1688 mm *Including built-in wheels
Depth 1075 mm
Weight
640 kg


Sterilization Chamber

Total Volume
142 L (5.01 cubic feet)
Dimensions (W x H x D)
425 mm x 425 mm x 790 mm
Shelf Strength
35 kg
Shelf Dimensions (W x D)
Upper: 415 mm x 775 mm
Lower: 410 mm x 775mm


STERILIZATION MODE OVERVIEW

SURFACE MODE

35min

ENDO MODE

50min

STANDARD MODE

62min

ADVANCE MODE

68min

Quick Surface Sterilization Cycle

  • General metal/non-metal, non-lumened devices requiring surface
    sterilization, or sterilization of mated stainless steel or titanium surfaces

  • Non lumen rigid, semi-rigid endoscopes





Efficient Flexible Endoscope Sterilization Cycle

  • Three(3) Single channel Flexible endoscopes with :
    Internal diameter of 1mm or larger and a length of 850mm or shorter

  • Two(2) Dual channel Flexible endoscope with :
    Internal diameter of 1mm or larger and a length for one channel of
    998mm or shorter and the other channel of 850mm or shorter
Conventional Flexible Lumen Sterilization Cycle with Extra Loads

  • Ten(10) Polyethylene and Teflon lumen instruments with :
    Internal diameter of 1 mm or larger and a length of 1,200 mm or shorter

  • One(1) Long Polyethylene and Teflon lumen instruments with :
    Internal diameter of 1mm or larger and a length of 4,000mm or shorter


Versatile Flexible/Rigid Sterilization Cycle with Extra Loads

  • Ten(10) Stainless steel lumen instruments with :
    Internal diameter of 1 mm or larger and a length of 500 mm or shorter

  • Five(5) Stainless steel lumeninstruments with :
    Internal diameter of 0.77 mm or larger and a length of 500 mm or shorter

One shelf of the chamber can be used in SURFACE mode.
Max. loading weight : 7.4kg (16.3lbs)
Up to three shelves of the chamber may be used in ENDO mode.
*Only Flexible Endoscopes with accessories may be loaded in ENDO mode
Up to two shelves of the chamber may be used in STANDARD mode.
Max. loading weight : 14.8kg (32.6lbs)
Up to two shelves of the chamber may be used in ADVANCED mode.
Max. loading weight : 14.8kg (32.6lbs)
* All medical devices to be sterilized must be washed and dried completely prior to sterilization in HMTS system.